Back to Search
Journal ArticleOpen Access

Acceptance Tests of more than 10 000 Photomultiplier Tubes for the multi-PMT Digital Optical Modules of the IceCube Upgrade

Author Affiliations
Loyola University Chicago, Deutsches Elektronen-Synchrotron DESY, University of Canterbury, University of Wisconsin–Madison, ...
Published InJournal of Instrumentation
Year2024
Citations3

Abstract

Abstract More than 10000 photomultiplier tubes (PMTs) with a diameter of 80 mm will be installed in multi-PMT Digital Optical Modules (mDOMs) of the IceCube Upgrade. These have been tested and pre-calibrated at two sites. A throughput of more than 1000 PMTs per week with both sites was achieved with a modular design of the testing facilities and highly automated testing procedures. The testing facilities can easily be adapted to other PMTs, such that they can, e.g., be re-used for testing the PMTs for IceCube-Gen2. Single photoelectron response, high voltage dependence, time resolution, prepulse, late pulse, afterpulse probabilities, and dark rates were measured for each PMT. We describe the design of the testing facilities, the testing procedures, and the results…
View at Publisher

BORR does not host full-text PDFs. The button above takes you to the original publisher.