Back to Search
Journal ArticleOpen Access

Speciation of arsenic by ion chromatography inductively coupled plasma mass spectrometry using ammonium eluents

Author Affiliations
University of South Australia
Published InJournal of Separation Science
Year2006
Citations36

Abstract

A method based on ion chromatography (IC) and inductively coupled plasma MS (ICP-MS) was developed for the speciation of arsenic in water and soil extracts. An anion-exchange column (G3154A/101) was used to separate As(III), As(V), dimethylarsinic acid (DMA), and monomethylarsonic acid (MMA) with excellent resolution. Various ammonium salts, including NH4H2PO4, (NH4)2HPO4, (NH4)2CO3, and NH4HCO3, were examined as eluents to reduce matrix interference from chloride and to solve clogging problems. The best arsenic speciation was obtained within 9 min with excellent resolution and without interference from high chloride concentrations using an eluent containing 7.5 mM (NH4)2HPO4 at pH 7.9. The detection limits for the target arsenic species ranged from 0.1 to 0.4 microg/L with direct injection of sample without matrix elimination.…
View at Publisher

BORR does not host full-text PDFs. The button above takes you to the original publisher.