Magnetic resonance imaging in individuals with cardiovascular implantable electronic devices
Authors
Author Affiliations
Rappaport Family Institute for Research in the Medical Sciences, Technion – Israel Institute of Technology, Dhaka International University, University Hospital of Zurich, ...
Published InEP Europace
Year2008
Citations255
Abstract
Magnetic resonance (MR) imaging has unparalleled soft-tissue imaging capabilities. The presence of devices such as pacemakers and implantable cardioverter-defibrillators (ICDs), however, was historically considered a contraindication to MR imaging. We summarize the potential hazards of the device-MR environment interaction, and present updated information regarding in vitro and in vivo experiments suggesting that certain pacemaker and ICD systems may indeed be MR-safe. Recent reports on several hundred patients with implantable pacemakers and ICDs who underwent MR scan safely indicate that, under certain conditions, individuals with these implanted systems may benefit from MR imaging. We believe that, on a case-by-case basis, the diagnostic benefit from MR imaging outweighs the presumed risks for some pacemaker and ICD patients. Thus for some patients, the…